Reflective SpO2 Module 'PPG-2R-880/940'
Suitable for the development of reflective PPG products! Outputs analog signals of R and IR.
The "PPG-2R-880/940" is a reflective SpO2 performance tester designed for research and development purposes by the manufacturer. It works in conjunction with the AECG100 main console or SECG5.0 AIO to test heart rate and blood oxygen saturation functions. The amplitude and frequency of R/IR in PPG, as well as the rate of change of the respiratory signal, can be individually adjusted, allowing for more thorough validation. 【Features】 ■ Capable of outputting R and IR analog signals that simulate SpO2 signals ■ Adjustable SpO2 parameters such as AC value, DC value, and PI value of R/IR ■ Respiratory variability simulation - allows for more comprehensive testing with add-on signals ■ Noise settings with adjustable frequency and amplitude can be output together with PPG signals *For more details, please download the PDF or feel free to contact us.
basic information
【Specifications (Partial)】 ■Heart Rate ・Setting Range: 10–300 BPM ・Minimum Step Size: 1 BPM ・Accuracy: ±1 BPM ■LED DC Level Reference Output ・Setting Range: 100–3000 mV ・Minimum Step Size: 1 mV ■LED AC Level Reference Output ・Setting Range: 0.75–30 mV ・Minimum Step Size: 0.01 mV *For more details, please download the PDF or feel free to contact us.
Price range
Delivery Time
Applications/Examples of results
For more details, please download the PDF or feel free to contact us.
catalog(1)
Download All CatalogsRecommended products
Distributors
We are a trading company that has been supporting manufacturing in the IoT era for 70 years. Our domestic locations include our headquarters in Kyoto (Minami-ku, Kyoto City), a sales office in Nagoya (Meito-ku, Nagoya City), a sales office in Shiga (Ritto City, Shiga Prefecture), a branch office in Akashi (Kakogun, Hyogo Prefecture), and a sales office in Okayama (Kita-ku, Okayama City), as well as two overseas locations in Malaysia and the Philippines.